BEGINNER ROUTE · MOVE ANYWHERE WITHOUT LOSING YOUR PLACE

Choose a lesson.

CHAPTER 01 OF 14See the whole scanner

Start with the patient, table, nested scanner hardware, and the order in which an MRI measurement happens.

Interactive MRI physics00 — 09

You don’t photograph a body.
You encode it.

Start with RF excitation and relaxation, then follow gradients as they turn position into frequency and phase—from body and local coils to an oblique trajectory through k-space and an image reconstructed line by line.

3.0 T
42.58 MHz for each tesla
Gradient echo (GRE)
ƒEquation ?meaning · units · live what-if · why care Move / zoom / explain 3Ddrag to orbit · use + / reset / − · tap an object to identify it Change a controloutputs and cause/effect notes update together
Interactive 3D scanner
GUIDED 3D MODEL · VIEWS 03–05 FOLLOW EVENT ORDER Patient + table

The table carries the patient into the fixed bore. Table motion changes which anatomy is near isocenter; it does not move through k-space and does not electronically select a slice.

SOLID physical hardware GLOW / ARROWS invisible field cue PLANE / CLOCKS / BOXES selected region or calculated data cue

TEACHING VIEW ONLY · THE INSTALLED SCANNER DOES NOT PULL APART
nested as installedpulled apart to teach

At 0%, the coil layers sit where they are installed: concentrically around the bore.

Camera preset · patient positioning

Drag to orbit · pinch or wheel to zoom · tap a label or object to explain it

START SIMPLE · ONE STORED SAMPLE MIXES SIGNAL FROM THE WHOLE EXCITED REGION

S(k) = ∫ ρeff(r) e−i2π k·r dr

In plain words: treat the excited body as many tiny, equal-size regions. Each region contributes some receive-coil voltage. Gradients give those contributions a calculable phase—where they sit around one repeating cycle. At address k, the receiver adds every region and stores one pair of signed numbers, I and Q, called S(k). That pair describes the whole excited region; it is not one image pixel. Tap the equation for every symbol, unit, zero case, and a live numerical example.

SEARCHABLE MRI DICTIONARY · 123 / 123 TERMS HAVE THE COMPLETE BEGINNER EXPLANATION

Type the term as you saw it—or describe what confused you.

Every result starts without assumed MRI knowledge, then lets you go deeper into a labeled visual, scanner behavior, clinical consequence, numerical example, and connected terms.

  1. 01Age-10 meaning
  2. 02Name + symbol decoded
  3. 03Physical, calculated, or displayed?
  4. 04Exact unit or no unit
  5. 05Why MRI needs it
  6. 06What fails without it
  7. 07What more / less / negative does

BEGINNER-FIRST LEARNING · CHOOSE HOW FAR TO ZOOM

Start with one plain idea. Add detail only when you want it.

The same three-level control appears as a short entry card at every major topic. It changes the explanation layer—not the MRI physics or any simulated parameter.

START SIMPLEEach chapter now begins with one plain-language mental model. Use “Zoom in” on any card when you are ready.

THE WHOLE MRI EXAM · NOT ONLY RF, GRADIENTS, OR K-SPACE

Follow the patient, hardware, signal, and data from room entry to finished images.

An MRI scanner is a coordinated system: safety screening, a continuously energized main magnet, a moving patient table, shim and gradient coils, RF transmit/receive, monitoring, sequence control, digitization, corrections, reconstruction, storage, and display.

Side cutaway · fixed scanner + moving patient supportSTEP 01 · SCREEN + PREPARE
Interactive complete MRI scanner side cutaway and patient table movement A patient table moves through a fixed magnet bore toward isocenter while highlighted subsystems show screening, magnetic field, shimming, RF and gradient sequence operation, reception, and reconstruction. FIXED MAGNET + CRYOSTAT B₀ remains energized between exams ISOCENTERfixed scanner origin B₀ DIRECTION SHIM COILS / CALIBRATION VOLUME RF EXCITATION + SWITCHED GRADIENT FIELDS TINY RF SIGNAL → LOCAL COIL FIXED TABLE TRACK / DRIVE SELECTED ANATOMICAL LANDMARK ROOM LASER / LANDMARK REFERENCE CONTROLLED MRI AREAscreen implants + objectsposition coils + cableshearing + communication SEQUENCECONTROLLER timed commands CORRECT · COMBINE · RECONSTRUCT I/Q samples TABLE MOVES PATIENT MECHANICALLY · SCANNER HOUSING + ISOCENTER STAY FIXED
ROOM + PATIENT SYSTEMS

Preparation, table, coils, communication, and monitoring

Technologists screen the patient and every entering object, position anatomy and coils, choose a landmark, provide hearing protection and an alarm device, and use MR-conditional monitoring or gating when required.

MAGNET + CRYOGENIC SYSTEM

Main field, cryostat, shielding, cooling, and quench protection

The main magnet supplies B₀ continuously. The cryostat thermally supports the superconducting system; site infrastructure, shielding, and emergency procedures manage fields and rare abnormal events.

FIELD PREPARATION

Localizers, shimming, frequency adjustment, and calibration

Fast survey images establish geometry. Field mapping and shim adjustments improve uniformity; reference scans can estimate coil sensitivity, center frequency, transmit behavior, and reconstruction corrections.

SEQUENCE + POWER HARDWARE

Precisely timed RF, gradients, receive switching, and ADC

The controller schedules RF synthesizers/amplifiers, gradient amplifiers, transmit/receive protection, receiver bandwidth, and ADC sampling. Timing—not a single component—defines the acquisition.

COMPUTE + RECONSTRUCTION

Corrections, channel combination, Fourier encoding, and image formation

Raw multi-channel I/Q data may be corrected for sampling and hardware behavior, calibrated, reconstructed, combined across coils, filtered, scaled, and packaged with geometry and protocol metadata.

DISPLAY + CLINICAL WORKFLOW

Series, image magnitude/phase, measurements, storage, and interpretation

The console and downstream systems show reconstructed images with orientation, scaling, annotations, and metadata. A displayed gray level is the end of a long weighted chain—not a direct photograph or universal tissue unit.

ρ EXPLAINED WITHOUT HIDDEN FACTORS · SOURCE → ECHO → COIL → IMAGE

What “rho” actually means—and what more or less of it changes.

Textbooks often reuse ρ for both proton density and the already-weighted signal distribution. This lab keeps those quantities separate, names every factor in the teaching product, and states the reference behind every “relative” number.

“MR-VISIBLE” MEANS

Mobile ¹H signal that can join a detectable echo.

Most clinical proton MRI signal comes from hydrogen nuclei in mobile water and fat. Hydrogen locked in very rigid material can lose transverse coherence before the receiver can sample it; air has very few hydrogen nuclei. “Visible” never means visible light.

“RELATIVE” MEANS

Compared with one declared reference—not an absolute proton count.

Here, an equal-size reference voxel with ρH = 1.00, complete recovery, no T₂* loss, a 90° excitation, and receive sensitivity 1.00 has ρeff = 1.00. Scanner gain and display windowing can rescale all image numbers, so there is no universal brightness or volt value.

“ON THE IMAGE” MEANS

The reconstructed voxel at that physical position.

One k-space measurement is not a pixel. After all complex samples are reconstructed, the ideal local complex value is proportional to ρeff. A magnitude image displays its size, usually after coil combination, scaling, filtering, and window/level.

One equal-size voxel through five named multipliersREFERENCE-NORMALIZED TEACHING MODEL
Interactive source-to-received-signal factor model Hydrogen source density is multiplied by T1 recovery, T2-star survival, excitation, and receive-coil sensitivity to form an effective local signal contribution. 01 · LOCAL SOURCE ρH = 0.80 equal voxel · relative to reference 02 · SEQUENCE + COIL FACTORS T₁ RECOVERY0.632 × T₂* SURVIVAL0.607 × EXCITATION1.000 × RECEIVE COIL1.000 03 · EFFECTIVE LOCAL SIGNAL AT THE ECHO ρeff = ρH × RT1 × DT2* × Esinα × Crx 0.307 × reference IDEAL VOXEL 0.307 AT k = 0 · this voxel contributes a positive complex arrow of length ρeff × voxel volume. AT k ≠ 0 · the arrow keeps that length but rotates by −2πk·r; other voxels can reinforce or cancel it. AFTER RECONSTRUCTION · the local complex value returns to this position in the ideal fully sampled model.
WHAT THIS SIMPLE PRODUCT INCLUDES

Local mobile-¹H source, one T₁ recovery term, one T₂* survival term, ideal sin α excitation, and one relative receive-sensitivity number.

WHAT REAL MRI MAY ALSO WEIGHT

Flow/inflow, diffusion gradients, magnetization transfer, chemical exchange, contrast agents, fat/water phase, B₀ and B₁ nonuniformity, motion, multi-echo history, receive-channel combination, filters, noise, gain, and display window/level. These are named here rather than hidden inside “other factors,” but they are deliberately held out of this five-factor lab.

CLINICAL READING RULE

Brighter does not automatically mean “more protons.” First ask whether sequence timing, excitation, coil position, pathology, reconstruction, or display scaling also changed.

COORDINATE PRIMER · FIX THE NAMES BEFORE ENCODING

X, Y, and Z name fixed hardware.
Read, phase, and slice name jobs.

An axial example often pairs read with Gx, phase with Gy, and slice with Gz—but that pairing is not a law. Rotate the prescribed image plane and the scanner synthesizes each logical job by firing two or three physical gradient coils together.

WHY CARE · Confusing these naming systems can make an oblique image plane, artifact direction, or amplifier limit look wrong.
01 · CHOOSE IMAGE PLANE
02 · CHOOSE LOGICAL JOB
LOGICAL READ REQUEST +40.0 mT/m one requested vector magnitude · mT/m is field slope
CLICKABLE VECTOR RELATION

[Gx Gy Gz]ᵀ = [+1.000 0.000 0.000]ᵀ × +40.0 mT/m

Each coefficient is a unitless direction cosine. Superscript ᵀ means “write this row as a column vector” (transpose)—it does not mean tesla here. Multiplying by +40.0 mT/m gives a real physical-coil command.
Gxphysical X coil+40.0 mT/m
+1.000 unitless× logical request
Gyphysical Y coil0.0 mT/m
0.000 unitless× logical request
Gzphysical Z coil0.0 mT/m
0.000 unitless× logical request
VECTOR SUM Gx alone points along logical read. The bars are simultaneous amplifier commands, not three sequential encoding events.
MODEL BOUNDARY

The double-oblique basis is an illustrative orthonormal coordinate frame. A scanner calculates direction cosines from the prescribed patient/image orientation and applies calibration, ramp, slew-rate, duty-cycle, peripheral-nerve-stimulation, and hardware limits not solved in this compact primer.

00 / WHERE THE MAIN FIELD COMES FROM

Cold wire carries current.
The current creates B₀.

A power supply first pushes electric current through many turns of superconducting wire. Those turns act together like one long solenoid: their magnetic fields add through the bore and return outside the magnet. After ramp-up, a closed superconducting path can keep that current circulating in persistent mode.

Three-dimensional superconducting MRI magnet and magnetic-field model
RAMP SUPPLY CONNECTED PROBE · +2.95 T · +29,500 G Yellow arrowheads follow the purple wire through all 52 displayed turns. They show conventional-current direction, not electrons; electron drift is opposite. Mint arrows show the resulting +Z field direction.

Drag to orbit · pinch, wheel, or use + / ↺ / − to zoom · tap any object · field lines are a map, not physical tubes

Live superconducting-magnet model
IDEALIZED SOLENOID · NOT SERVICE CONTROLS
  1. 01
    Ramp the current

    A controlled DC power supply applies voltage so current rises in the cold winding. “Current” means net electric charge crossing a wire section each second, measured in amperes. Yellow arrows follow the conductor in the conventional positive-charge direction; the wire does not move, and electrons do not jump between turns.

  2. 02
    Every turn contributes

    Each loop makes a magnetic field. Inside a long coil, neighbouring loop fields point mainly the same way and add; more amperes or more turns per metre gives a larger central field.

  3. 03
    Close the persistent path

    Once the target current is reached, an internal superconducting connection completes a very-low-resistance loop. The ramp supply can be disconnected while the established current continues.

  4. 04
    B₀ stays on

    The stable main field exists between scans. RF and gradient pulses switch during imaging; the superconducting main-magnet current normally does not pulse on and off for each patient.

CLICKABLE LONG-SOLENOID ESTIMATE · AT THE CENTER

Bcenter ≈ μ0nI = μ0(N/L)I

4π × 10⁻⁷ T·m/A × 4,000 turns/m × 600 A ≈ 3.02 T
600 A 0 A · no current600 A teaching value800 A

At 600 A and the selected winding density, the ideal long-solenoid estimate is 3.02 T. Raising current strengthens B₀ in direct proportion. Yellow arrowheads stay on the helical wire and mark conventional-current direction; they are not electrons and their display speed is not a measurement.

4,000 turns/m 1,000 turns/m · 16 shownfixed 4.0 m length6,000 turns/m · 76 shown

The coil length stays 4.0 m. The 3D winding shows 52 spaced turns to represent 16,000 effective turns. Move the slider: higher n visibly adds turns and closes their gaps; the drawing remains a compressed sample rather than literal manufacturer geometry.

10 drawn loops 4 · uncluttered10 · balanced view16 · denser drawing

This changes only how many direction-map curves the 3D lesson draws. It does not change current, B₀, tesla, gauss, fringe field, stored energy, or any clinical image.

0.00 m −3 mcenter+3 m
0.00 m axisnear windingoutside
CURRENT × TURN DENSITY2.40 MA-turn/msource strength for this estimate
IDEAL CENTER B+3.02 T+30,159 G
NUMERIC PROBE B+2.95 T+29,500 G
YOU CHANGED CURRENT / WINDING DESIGN

More ampere-turns make a stronger main field.

At the center, increasing I or n increases the field almost linearly in this long-solenoid estimate. Clinically, a different B₀ changes the ¹H carrier frequency and many scanner/tissue behaviors; it is a magnet design and site platform, not a routine sequence knob.

WHAT PHYSICALLY MOVES?

Charge carriers have a tiny average drift inside the continuous wire. In a metal, negative electrons drift opposite the defined conventional-current arrow. They do not leap across the space between neighbouring windings; the wire, cryostat, and patient do not circulate.

WHAT ARE FIELD LINES?

Drawing lines connect the field direction at many points. They close into loops and crowd where the drawing represents stronger field. They are not strings, rays, current paths, or proton tracks.

WHY TESLA AND GAUSS?

Both measure magnetic flux density: 1 tesla = 10,000 gauss. Therefore 3 T = 30,000 G. Clinical MRI names B₀ in tesla; gauss is common in fringe-field discussions.

WHY SUPERCONDUCT?

Ordinary resistance would turn sustained high current into continuous heat. Below its critical conditions, the magnet conductor can support persistent current with extremely small decay; cooling and protection remain essential.

3D FIELD BUILD-UP · MOVING CHARGE → CURRENT → LOOP → SOLENOID

One conductor makes a circular field.
Many loop fields add.

Blue moving markers now represent groups of negative charge crossing a counting plane; increasing current shows more marker crossings and a stronger calculated field. They are not a literal census of electrons in the wire. Because electrons are negative, their average drift is opposite the yellow conventional-current arrow. Build from a straight segment to one loop, then watch the field shape lengthen as aligned turns are added.

Three-dimensional conductor, current-loop, and solenoid field-summation model
01 · STRAIGHT CONDUCTOR B circles the wire · reference |B| 2.67 µT Yellow I points +Z. The blue e⁻ tracer drifts −Z. Curled field arrows follow the right-hand rule for conventional current.

Drag to orbit · pinch, wheel, or use + / ↺ / − · tap an object · animation pace and line count are teaching choices

Build the field one geometry at a time
MICROTESLA TEACHING SCALE · NOT MRI-MAGNET SERVICE DATA
LIVE CHARGE-FLOW COUNTER · SYNCHRONIZED WITH THE 3D MODEL

Count charge crossing a wire section—not electrons merely sitting in the wire.

The side view is kept two-dimensional because “crosses this plane during one second” is a measured rate. The rotatable model beside it shows the same conductor and circular field in 3D.

CHARGE PER SECOND4.00 C/sthis equals 4.00 amperes
ELECTRON-CHARGE EQUIVALENTS2.50 × 10¹⁹ /scharge amount divided by |e|
VISIBLE MOVING PACKETS8 of 16compressed teaching markers, not eight electrons
ONE MARKER REPRESENTS3.12 × 10¹⁸ /sonly in this animation mapping
CLICKABLE RATE LAW · WHY “MORE ELECTRONS” NEEDS A TIME WINDOW

I = ΔQ / Δt = Ne|e| / Δt

Ampere measures charge crossing per second. Stationary charge gives no steady current-created B; current through the chosen geometry sets the calculated field.

At 4.00 A, 4.00 coulombs of charge cross the chosen section each second—equivalent to about 2.50 × 10¹⁹ elementary electron charges per second. The conductor already contains vastly more carriers; this control changes net flow rate, not the mere existence of electrons.

4.0 A · 4.0 C/s 0 A · carriers, no net flow4 A = 4 C/s8 A = 8 C/s

More charge crossing per second strengthens every B contribution in direct proportion. The number of bright blue packets is a compressed rate display; their visible speed is deliberately not a microscopic drift-speed measurement.

8 loops 1 loop8 visible contributors12 loops

Moving this control opens the many-loop stage. More equally oriented loops add more same-direction axial field at the center; it does not mean the 3D drawing represents a real MRI winding count.

CONVENTIONAL CURRENTI points +Zdirection assigned to positive charge
AVERAGE ELECTRON DRIFTe⁻ drifts −Znegative carrier direction is opposite I
ACTIVE CONTRIBUTORSone straight conductorfield circles the wire
CALCULATED REFERENCE FIELD2.67 µT0.30 m from an ideal long straight wire
COIL / FIELD SHAPEstraight-wire circular fieldfield direction circles locally around the Z-directed conductor
NET CHARGE-FLOW RATE2.50 × 10¹⁹ e⁻ charges/s4.00 C/s; not total electrons stored in the wire
WHAT THE 3D ARROWS ARE ADDING

Vector addition, not field-line counting

1Every short piece of current makes a small field vector at the chosen observation point.
2At the exact center of one circular loop, every ideal short-segment contribution points along the same axis and reinforces.
3Straight-wire stage: the rings show B direction at several distances; they are not separate extra fields.

QUANTITATIVE AXIAL CROSS-SECTION · SAME CURRENT AND WINDING CONTROLS

Color tells how much field.
Arrows tell which way it points.

The 3D loops above are useful direction guides, but their spacing is not a measurement. This map calculates a field value at a grid of positions through the coil. Drag the red probe directly: moving it changes only the observation coordinate, never the magnet.

Interactive finite-solenoid magnetic-field map.
CALCULATING FINITE-COIL FIELD GRID…

FINGER / POINTER: drag anywhere in the plot · KEYBOARD: focus the plot, then use arrow keys · the ordinary z and r sliders remain equivalent accessible controls

INTERACTIVE 3D · BODY → WATER → ¹H NUCLEUS → B₀ → RF → ENCODING → RECEIVE

Zoom from the patient to the nuclear physics.

Use the numbered journey in order. Every solid, sphere, arrow, field sheet, and ring can be tapped for its exact meaning. The jumps between body, tissue, atom, and nucleus are separate scale models—nothing literally swells inside the patient.

Three-dimensional body-to-hydrogen MRI physics journey
01 / 11 · BODY SCALE LOCATE ONE TISSUE REGION The yellow marker identifies the body region that later scale models magnify. It is not an MRI signal or a k-space address.
BODY · m TISSUE · mm MOLECULE · nm NUCLEUS · fm ENCODING · cm COIL SIGNAL · µV

Drag to orbit · pinch, wheel, or use + / ↺ / − · tap an object to identify it · model sizes are not one continuous scale

Live multiscale physics model
SLOWED + SCHEMATIC · VALUES PRINTED SEPARATELY
OPEN THE EXACT MEANING
SCALE MODEL 01 · METRES

First choose where inside the body the signal comes from.

The scanner surrounds the patient with B₀, gradient, and RF hardware. The yellow target marks one tiny tissue region inside the head. The next buttons magnify that region conceptually; the tissue does not move out of the body.

WHAT EXISTS PHYSICALLY?

Patient, table, scanner bore, tissue water, fat, and many other materials.

WHAT DOES THE DRAWING INVENT?

A glowing target and connector make one future region easy to follow.

WHY SHOULD I CARE?

MRI never receives a ready-made picture from one location; it must excite and encode signals distributed through the body.

3.0 T 0.5 T3 T7 T

At 3.0 T, ¹H phase precesses at 127.73 million cycles per second. The 3D arrow is slowed enormously; its animation speed is only a trend cue.

20.0 mT/m 0 · no slope20 mT/m40 mT/m

The selected stage assigns this field slope to physical Z for slice selection, Y for phase encoding, or X for readout. Gradient current changes precession rate with position; it does not push atoms along that axis.

2.0 kHz 0.5kHz6.0
0 mm −80mm+80

At 20.0 mT/m, a 2.0 kHz RF band selects an ideal 2.35 mm slab centered at z = 0 mm. Moving RF center frequency moves the slab; it does not move atoms.

¹H CARRIER AT ISOCENTER127.732 MHzcycles of transverse phase per second through time
SELECTED SLICE2.35 mm at z 0 mmRF center offset 0.00 kHz
PHASE LOBE · 0.80 msky +681.2 m⁻¹149.9 angle cycles across 220 mm
READOUT · Gx + ADCdkx/dt +0.852 × 10⁶ m⁻¹/sone dwell stores one whole-slice I + iQ sum
ANIMATION TRUTHMHz slowed to visible motionarrow direction and causal trends only; never timing calibration
CLICKABLE · FIELD → PRECESSION RATE

f0 = γ̄B0

CLICKABLE · RF BAND + GRADIENT → THICKNESS

Δz = BWRF / (γ̄|Gslice|)

CLICKABLE · READ GRADIENT → k-SPACE SPEED

dk/dt = γ̄G(t)

CLICKABLE · CHANGING FLUX → COIL VOLTAGE

vcoil(t) = −dΦ(t)/dt

THE NEXT SCALE CHANGEWatch repeated I + iQ samples fill k-space and become an image →The 3D journey ends at measured coil voltage. The reconstruction lab starts with the stored data—without pretending that a finished photograph was hiding underneath.

01 / FROM FIELD TO POSITION

A controlled tilt in the magnetic field.

A gradient coil adds a small, nearly linear variation to the much larger main field. Proton frequency now carries an address: where a spin sits determines how fast its phase turns.

B₀ − ΔB field strength → B₀ + ΔB
x = +8.0 cm
Live field model
¹H at B₀ = 3.0 T
LOCAL FIELD

Bz(r,t) = B0 + G(t) · r

3.0 T · 127.73 MHz 0.5 Tclinical 1.5 / 3 T7.0 T
25.0 mT/m −400+40 mT/m
+8.0 cm −12 cmisocenter+12 cm
ΔB+2.00 mT
Δf = γ̄ΔB+85.15 kHz
local f127.825 MHz
CARRIER LAW

f0 = γ̄B0

42.58 MHz/T × 3.0 T = 127.73 MHz
B₀ CHANGESRF carrier127.73 MHz
B₀ DOES NOT SETGradient offset+85.15 kHz
B₀ DOES NOT SETOn-resonance flip92.0°
B₀ DOES NOT SETk-space speed1.064 × 10⁶ m⁻¹/s

At fixed G, position, B₁⁺, and pulse duration, changing B₀ retunes the MHz carrier. The gradient offset, ideal flip angle, and dk/dt stay unchanged.

Scope: this control retunes the field/RF frequency model. The TE/TR tissue constants remain the explicitly stated illustrative 3 T model; real coil fields, SAR, and relaxation are field-dependent.

i The gradients are tiny beside B₀, but they are switched quickly and precisely. The audible knocking in MRI is gradient hardware moving under Lorentz force.

BEGINNER GAME / THREE HIDDEN WATER BUCKETS

You are the operator.
Find water with fields.

Three water buckets are hidden inside the bore. From the next room, choose a gradient vector, tune a narrow RF frequency offset, send a short probe, and listen for the echo. Find each bucket’s X, Y, and Z coordinate—then see why a frequency under one gradient selects a plane rather than a unique point.

WATER FINDER / CONTROL ROOM¹H · THREE HIDDEN OBJECTS · TEACHING GAME
Three-dimensional hidden water bucket game
SCANNER ROOMTHE OPERATOR IS OUTSIDE THIS ROOM
READY · BUCKETS HIDDEN Gx +12.0 mT/m · plane x = 0.0 cm The translucent square is every position that has the tuned frequency under the current gradient. It is a plane, not a flashlight beam.

Drag to orbit · solid ring = bore · blue plane = same-frequency locations · buckets reveal only after detection

OPERATOR NOTEBOOK · MEASURED COORDINATES

Match peak height across X, Y, and Z.

water volume / signal tagX coordinateY coordinateZ coordinatestatus
0.6 L · small peak???0 / 3 axes
1.0 L · medium peak???0 / 3 axes
1.4 L · large peak???0 / 3 axes
0 OF 9 COORDINATES FOUND One frequency measurement gives one projection—not a full 3D address.

With Gx on, frequency tells X but says nothing about Y or Z. Rotate the gradient job and repeat. Real MRI performs a much richer, systematic version with many phase and frequency encodings.

1

No gradient: no position clue

With G = 0, all three identical-water resonances sit near the same f₀. Frequency can say “water signal exists,” but not where it came from.

2

One gradient: one projection

Δf = γ̄G·r. A measured frequency gives position along G, while an entire perpendicular plane has that same frequency.

3

More encodings: reconstruct location

Real imaging collects many known phase/frequency patterns. Reconstruction solves the mixed whole-object voltages for a spatial image; an operator does not hunt voxel by voxel.

GAME MODEL BOUNDARY

These buckets contain ideal identical water but have different volumes, so the game uses peak height as a tracking label while assuming uniform transmit/receive sensitivity, no relaxation difference, no noise, and a narrow Gaussian slice profile. Real peak amplitude also changes with coil sensitivity, loading, timing, voxel volume, motion, and noise; clinical MRI uses systematic spatial encoding and reconstruction, not manual frequency hunting or bucket-volume tags.

02 / RF TRANSMIT, RELAXATION & RECEIVE

Excite broadly.
Listen locally.

The transmit coil creates the rotating B₁⁺ field that tips magnetization. After excitation, the scanner stops transmitting and receive coils detect the tiny voltage induced by precessing transverse magnetization. A common 1.5 T / 3 T workflow is body-coil transmit with a close local array receiving.

RF COIL CHAIN / ¹H AT 3.0 TTX / RX SWITCHING · IDEALIZED FIELD MODEL

RF WAVEFORM DECODER · THREE DIFFERENT SHAPES, THREE DIFFERENT JOBS

Frequency is horizontal spacing. B₁⁺ amplitude is height. Pulse duration is width.

The plots share the live controls below. They are separated because “a faster wave,” “a stronger pulse,” and “a larger received signal” do not mean the same thing.
Change B₀ → retune the carrier in this comparison
T means tesla, the unit of magnetic-field strength.
RF carrier, transmit envelope, and received-signal decoder The top row shows one component of the RF magnetic field over 80 nanoseconds. The middle row shows the slower transmit pulse envelope over 3 milliseconds. The bottom row shows received echo magnitude after carrier removal. 01 · CARRIER ZOOM one B₁ field component fixed window: 80 ns spacing = frequency height = B₁⁺ amplitude one cycle = 7.83 ns 0 ns 80 ns 02 · TRANSMIT ENVELOPE outline joining carrier peaks window: 0 to 3 ms height = field strength width = pulse duration 6.0 µT for 1.00 ms 0 ms 3 ms 03 · RECEIVED ENVELOPE echo voltage magnitude after carrier removal height = receive proxy not transmit B₁⁺ 1.00× geometry-only receive sensitivity receiver window opens later
One RF cycle

The plotted field goes from a positive peak, through zero and a negative peak, back to the next positive peak: one complete 360° oscillation. It is a cycle of the RF magnetic field—not a proton traveling in a circle.

Envelope

The slow outline connecting the peaks of thousands of fast carrier cycles. It is not a second radio wave and not a container; it summarizes how strongly the transmitter is driven over time.

Received magnitude

The size of the complex voltage after the receiver mathematically removes the MHz carrier. Phase is stored too, even though this row shows only non-negative magnitude.

MODEL BOUNDARY

This decoder uses a constant-frequency rectangular pulse. Real MRI can use shaped, adiabatic, multiband, or frequency-swept pulses, whose envelope and instantaneous frequency can both vary. The receive curve holds tissue, flip, TE, noise, loading, and reconstruction fixed so coil geometry can be isolated.

Use the B₁⁺, duration, distance, channel, coil-routing, and transmit/receive controls immediately below.
RECEIVER SIGNAL PATH · BEFORE AND AFTER DEMODULATION

How one fast RF coil voltage becomes I and Q.

Before demodulation there is one rapidly alternating voltage. The receiver compares it with two synchronized reference waves; low-pass filtering leaves two slower signed voltages that preserve the signal’s magnitude and phase.

Same signal · four viewsNORMALIZED PHASE +45°
Interactive RF demodulation into I and Q A fast received RF voltage is compared with cosine and quarter-cycle-shifted sine references. Low-pass outputs form a two-dimensional I/Q vector. BEFOREcoil voltage vRF(t)one fast waveform I MIXERcosine reference0° reference axis Q MIXERsine reference90° from I AFTER FILTERslow analog outputs MULTIPLY +LOW-PASS MULTIPLY +LOW-PASS I+0.707 Q+0.707 +I+Q φ 45° Wave spacing represents carrier frequency. This fixed teaching window shows 6 cycles—not 127.73 million cycles. I/Q axes are mathematical reference components. They are not physical scanner X/Y/Z directions.
MODEL BOUNDARY

The plot slows the carrier to six visible cycles and normalizes amplitude to A = 1. Real receiver architectures may digitize before or after analog mixing, use different Q sign conventions, and report relative digital units after gain. The two-component information is the same.

Three-dimensional RF transmit and receive model
TRANSMIT WINDOW · B₁⁺127.73 MHz
BODY COIL → B₁⁺ → SPINS α = 92.0° Local receive elements are detuned while the body coil transmits.

Drag to orbit · mint cage is body RF · coral loops are local elements

TRANSMIT FLIP ANGLE

α = γ ∫ B₁⁺(t) dt

RECEIVED VOLTAGE

vRx(t) ∝ −dΦM/dt

6.0 µT 2 µTRF field strength18 µT
1.00 ms 0.20 mspulse area3.00 ms
3.0 cm close · 1 cmreceive geometryfar · 12 cm
Visible local array elements
RF waveformpower amplifier
body coilB₁⁺ transmit
magnetizationprecessing spins
local arrayB₁⁻ receive
preampsADC / k-space
BODY TRANSMIT · LOCAL RECEIVE

The large built-in body coil drives a broad, comparatively uniform B₁⁺ field. During transmit the nearby receive array is actively detuned; during reception the transmitter is isolated and the local elements feed low-noise preamplifiers.

predicted rectangular-pulse flip92.0°
receive sensitivity proxy1.00× local
excitation coveragebroad / uniform
receive coveragelocal · 8 channels
transmit safety focuswhole-body SAR
switch stateRx array detuned

The sensitivity number is a normalized geometry teaching proxy, not a scanner specification. The shared B₀ control retunes the displayed ¹H carrier; it does not rescale this idealized coil geometry, B₁ field, SNR, relaxation, or SAR model. Actual values require coil- and patient-specific electromagnetic measurements.

LIVE 3D LOCAL-COIL RECEIVE JOURNEY · USES THE MODEL ABOVE

What a local coil physically captures—and what happens after.

Drag the stage scrubber with a finger or press Play. The 3D camera follows the signal from the patient to the final combined image.
STAGE 00 OF 07 · HARDWARE HANDOFF Stop transmitting before the tiny receiver listens.

The body coil finishes the high-power B₁⁺ pulse. The transmitter is isolated, the receive-only local loops come out of their protected detuned state, and their low-noise paths are connected.

WHAT CAUSES THIS?

The programmed pulse ends and the transmit/receive switching network changes electrical connections after a short recovery interval.

WHAT EXISTS NOW?

Excited transverse magnetization exists in the patient. The local array is now electrically able to respond to it.

MEASURED IN WHAT?

Switch timing is measured in seconds, commonly microseconds (µs). No image value has been measured yet.

WHY DO WE CARE?

The receive signal is tiny compared with transmit power. Isolation protects the preamplifier and prevents transmitter leakage from hiding early signal.

DO NOT CONFUSE IT WITH

The RF pulse did not travel into the local coil as an image. It prepared magnetization; reception is a later electromagnetic induction event.

00 · Tx → Rx transmit endsdrag / swipe through processingcombined image
1.00× 0.25×display speed only2.00×
PREAMPLIFIER VOLTAGE GAIN · TAP FOR UNITS AND CLIPPING

GdB = 20 log10(Vout/Vin)  ↔  Vout = Vin·10GdB/20

Gain scales signal and existing input noise; it does not create localization or improve input SNR.
40 dB · ×100 20 dB · ×10voltage scaling60 dB · ×1000
128 kHz 32 kHzsample rate / noise tradeoff256 kHz

CURRENT SETTINGSAt 40 dB, voltage is multiplied by 100 before digitization. A 128 kHz sample rate gives 7.81 µs between samples in this one-sample-per-dwell model.

IN PATIENTMxy(r,t)

Moving transverse magnetization distributed over position r.

AT COIL cvc(t) ∝ −dΦc/dt

One continuous signed RF voltage per element.

AFTER DEMODULATIONSc(t) = Ic(t) + iQc(t)

Two slower signed components per element.

AFTER ADCSc[n] at k[n]

Digital complex samples assigned to gradient-created addresses.

SIMPLE COMBINATION EXAMPLEIRSS(r) = √Σc|Ic(r)|²

Aligned coil images become one relative magnitude image; tap for limits.

WHAT THIS LIVE MODEL HOLDS FIXED

The 8.0 µV reference and 0.40 µV RMS noise at 128 kHz are illustrative, not specifications. Tissue amount, TE, relaxation, loading, tuning, cable loss, noise correlation, filters, gradients, and reconstruction settings are fixed unless named. The distance curve and √bandwidth noise rule show direction of change only. Real arrays require measured complex sensitivity maps and noise covariance; channel count alone does not promise an SNR gain.

Color code in 3D: solid boxes/loops are hardware · glowing curves and beads are invisible physical or electrical signals · grids and image cards are calculated data displays.
B₁⁺

Transmit field

RF power at the Larmor frequency rotates magnetization. Amplitude and pulse duration set flip angle; spatial B₁⁺ variation makes flip angle nonuniform.

B₁⁻

Receive sensitivity

Precessing transverse magnetization induces a tiny voltage. Close local elements couple strongly to nearby anatomy and admit less distant noise.

T/R

Isolation matters

Receive-only elements are detuned during the high-power transmit pulse. The transmit path is then isolated while low-noise preamplifiers listen for the echo.

TE / TR RELAXATION CLOCKILLUSTRATIVE 3 T TISSUE MODEL · 128 PHASE ENCODES
Predicted relative signalSPIN ECHO · TR 500 / TE 15 ms
fatgraywhiteCSF
Recovery before RF / decay before echoT₁ RECOVERY + T₂ DECAY
IDEAL SPIN-ECHO SIGNAL

S = ρH(1 − e−TR/T₁)e−TE/T₂

The 180° pulse refocuses static dephasing, so ideal echo amplitude follows T₂ rather than T₂*.
500 ms 200 msrecovery + scan time6000 ms
15 ms 5 mstransverse decay200 ms
90° fixed spin echo fixes 90°60°
WHITE MATTERρ 0.70 · T₁ 850 · T₂ 80 · T₂* 55 ms
CSF / FLUIDρ 1.00 · T₁ 4000 · T₂ 2000 · T₂* 400 ms
One repetition90° → 180° → ECHO → NEXT 90°
RFSIG 90° 180° TE 15 ms TR 500 ms
SHORT TR · SHORT TE

Short TR samples tissues before full longitudinal recovery, strengthening T₁ differences. Short TE limits T₂ decay, so the current spin echo is predominantly T₁ weighted.

white matter signal0.280
CSF / fluid signal0.114
pairwise contrast42.2%
dominant weightingT₁ weighted
2D scan-time proxy1:04 · 128 lines
echo fraction TE / TR3.0%

Signals use simplified steady-state equations and illustrative relaxation values near 3 T. Real contrast also depends on sequence details, RF profiles, echo trains, magnetization transfer, flow, diffusion, coils, reconstruction, pathology, and field strength.

03 / THREE ORTHOGONAL CONTROLS

Same physics.
Three jobs.

Read, phase, and slice are logical jobs attached to the prescribed image plane. The X/Y/Z labels in this first axial example name image coordinates—not a permanent promise that read = physical Gx, phase = physical Gy, and slice = physical Gz. For an oblique plane, the scanner combines all three fixed physical coil sets.

THE THREE TABS BELOW COMPARE ROLES; THEIR LEFT-TO-RIGHT ORDER IS NOT TIME. In the simplified 2D sequence taught here, RF plus the slice-select gradient first creates transverse signal in a slab. A later phase-encoding lobe creates a retained angle pattern. A read prephaser then prepares the start, and only afterward do the read gradient and ADC collect samples.

Three-dimensional readout, phase-encoding, and slice-selection comparison
READOUT · ADC OPEN Gread +20.0 mT/m Different x positions have different receive-frequency offsets while whole-object I/Q samples are stored.
low fhigh f

Drag to orbit · pinch, wheel, or use + / ↺ / − · tap an object to identify it

GX · ON DURING ADC

Position becomes frequency.

During signal readout, Gx makes spins at different x positions precess at different frequencies. Sampling through time walks continuously across one row of k-space.

FREQUENCY OFFSETΔf(x) = γ̄ Gx x
gradient actionContinuous plateau
moves throughkx within a line
resolution set by±kx,max
250 A 0 A · no added slope250 A teaching value400 A

With teaching efficiency η = 0.080 mT/m/A, 250 A produces 20.0 mT/m. During readout that increases frequency separation and moves farther through kx during a fixed time.

CURRENT → GREAD → FREQUENCY → KX

Follow what the current actually changes.

01AMPLIFIER COMMAND250 Aelectric charge-flow rate in one gradient circuit
02CALIBRATED FIELD SLOPE20.0 mT/mηI; Bz changes with logical position
03ACROSS 220 mm FOV187.3 kHz spreaddifference between the two edges while G is on
04AFTER FIXED 0.80 ms149.9 phase cyclesrelative angle turns accumulated edge-to-edge
05ENCODING CONSEQUENCEkx extent ±340.6 m⁻¹about 1.47 mm ideal read detail for the stated symmetric traversal

HELD FIXED: 220 mm FOV, 0.80 ms teaching interval, coil efficiency, RF bandwidth, and ideal sampling. Hardware current alone does not promise clinical resolution.

INTERACTIVE 3D · ONE 2D CARTESIAN REPETITION IN TRUE EVENT ORDER

Watch slice, phase, and readout become three different jobs.

The translucent oval is one digital volume. The violet slab is the part given transverse signal. Small white hands show the phase of spin packets inside that slab. They are not individual protons. Follow the numbered events; then rotate the prescribed plane and see the same logical jobs redistributed across physical Gx, Gy, and Gz.

01 · ORIENT THE PRESCRIPTION NO ENCODING GRADIENT YET longitudinal magnetization · no transverse receive signal
digital object volume selected slab RF transmit cue transverse-phase hands stored I + iQ sample

Drag to orbit · pinch, wheel, or use + / − to zoom · tap an object to identify it

01 · PRESCRIBED IMAGE PLANE

Axial example: logical read = physical X, logical phase = physical Y, and logical slice = physical Z. This simple one-to-one mapping changes when the plane rotates.

EVENT 01 / 07 · GEOMETRY PRESCRIBED

THE SCANNER DEFINES THREE LOGICAL DIRECTIONS

Read and phase lie inside the image plane. Slice points perpendicular to it. No RF pulse has tipped magnetization yet, so the white transverse-phase hands are hidden and the receiver has nothing useful to read.

LOGICAL GRADIENT JOBOFForientation is geometry, not a gradient pulse
RF TRANSMITOFFno excitation yet
ADC / RECEIVER SAMPLINGCLOSEDno voltage sample is stored
CALCULATED K-SPACE ADDRESSkx 0 · ky 0an address from gradient area—not a body location
ACTIVE LOGICAL VECTOR → FIXED PHYSICAL COILS No gradient command
Gx0.0 mT/m
Gy0.0 mT/m
Gz0.0 mT/m

The plane can be prescribed before any amplifier fires. When a logical gradient is requested, its direction cosines set the simultaneous Gx/Gy/Gz mixture.

WHY THE SLICE GRADIENT REVERSES AFTER RF

During a symmetric RF pulse, selected spins are tipped over a finite time while the slice gradient remains on. The gradient area after the pulse’s effective center leaves a position-dependent phase slope. A following opposite logical-slice area—about half the full selection plateau in this simplified case—cancels that slope. It does not undo excitation or select a second slice.

WHY SLICE?

Limit which slab creates transverse signal

The logical slice gradient makes resonance frequency vary along the plane normal. RF excites its chosen frequency band, so only the matching slab is tipped. Without slice selection, signal from the larger excited volume would overlap in a 2D image.

WHY PHASE?

Create an independent position-dependent angle pattern

A brief post-excitation gradient makes positions accumulate different transverse angles. The lobe then turns off, but the relative angles remain. Repeating a different signed area on later TRs supplies the independent ky measurements needed to separate positions along that direction.

WHY READOUT?

Measure many kx addresses while ADC is open

The read gradient creates position-dependent frequency offsets and moves the calculated address through kx. The ADC stores a new whole-slice I + iQ coefficient at each dwell. Without it, one repetition would not efficiently sample a complete k-space row.

MODEL BOUNDARY · The oval, points, phase hands, color bands, RF rings, and seven-address readout are an explanatory digital model—not anatomy, individual nuclei, a measured B field, or a clinical pulse-sequence prescription. The “about half-area” slice-rephasing statement assumes a symmetric RF envelope on a flat selection gradient and refers to gradient-created phase from the effective RF center. Real pulse shapes, gradient ramps, , refocusing pulses, , , , and manufacturer implementation can change the exact waveform. In an axial plane the logical slice job can be physical Gz; in an oblique plane its reverse lobe reverses the required Gx/Gy/Gz mixture, not necessarily Gz alone.

READOUT VS PHASE · EVENT-BY-EVENT

After RF creates transverse signal, set one ky address. Then sample across kx.

They are not two different kinds of magnetism. Both logical jobs use a gradient, and either gradient creates position-dependent frequency offsets while it is on. Their timing relative to the receiver is what makes their stored information different.

IMAGE PLANE · WHICH PHYSICAL COILS DO THE JOB?
In this axial example, logical read uses physical Gx and logical phase uses physical Gy.
ky = −2 Δk −3 Δksigned gradient area · not y position+3 Δk
0.5×changes animation only
EVENT 01 / 08 · BEFORE IN-PLANE ENCODING

RF CREATES TRANSVERSE SIGNAL

The RF pulse tips magnetization so a receive signal can exist. RF transmit is not readout: the receiver is protected and ADC is closed during excitation. Neither in-plane logical gradient job has encoded a k-space row yet.

READOUT MOTION THROUGH K-SPACE dkread/dt = γ̄ Gread(t)

A stronger read gradient crosses k-space faster. With dwell time fixed, that changes sample spacing and therefore FOV; with the acquisition prescription adjusted, it also affects bandwidth and distortion.

WHY CARE · This job sets read-direction sampling, bandwidth, chemical-shift displacement, and distortion behavior.
RETAINED PHASE PATTERN Δφ(yimage) = 2π (−2 Δk) yimage

After RF has created transverse magnetization, the signed area under the phase gradient creates a known phase-versus-position ramp and therefore sets ky. Changing that commanded area on successive TRs sets different ky addresses; it does not move to a literal y location in the patient.

WHY CARE · Phase steps strongly affect scan time, phase FOV, wrap, motion ghosts, and the direction of many artifacts.
SAME PHYSICS

Both cause Δf while on

A magnetic-field gradient changes local precession frequency. “Frequency encoding” and “phase encoding” describe how the sequence uses the accumulated effect, not two different gradient mechanisms.

DIFFERENT TIMING

ADC open versus ADC closed

Readout samples continuously while Gread is on. Phase encoding applies a lobe before sampling, closes it, and carries the retained phase ramp into the readout window.

PRACTICAL CONSEQUENCE

One row per repetition

Readout collects many kx points in one ADC window. Conventional 2D Cartesian imaging repeats the TR with a new phase area to cover many ky rows, so the phase loop often dominates scan time.

MODEL BOUNDARY · Events are separated here so each cause is visible. Real pulse sequences often overlap the read prephaser and phase-encode lobe, use tens to thousands of samples, include finite ramps and spoilers, and may collect multiple ky lines per TR with echo trains or segmented readouts.

OBLIQUE COORDINATE MIXERLOGICAL AXES → PHYSICAL GRADIENT AMPLIFIERS
Three-dimensional oblique coordinate model
logical read / phase / slice frame fixed physical Gx / Gy / Gz frame
LOGICAL READOUT GRADIENT +30.0 mT/m Gx +30.0 · Gy 0.0 · Gz 0.0 mT/m

Drag to orbit · plane normal is logical slice Z

COORDINATE TRANSFORM

Gphysical = R · Glogical

Columns of R are the physical directions of logical readout, phase, and slice. Because R is orthonormal, the three logical axes remain mutually perpendicular.
−90°physical X rotation+90°
−90°physical Y rotation+90°
−90°about slice normal+90°
+30.0 mT/m −70reverse at 0+70 mT/m
Rotation matrix R · physical rows by logical columns
physicalReadPhaseSlice
Gx+1.0000.0000.000
Gy0.000+1.0000.000
Gz0.0000.000+1.000
PHYSICAL AMPLIFIER COMMANDS±40 mT/m per-axis model
GX+30.0 mT/m
GY0.0 mT/m
GZ0.0 mT/m
AXES ALIGNED

In an axial prescription, logical readout aligns with the physical X gradient. Only the Gx amplifier is needed for this +30.0 mT/m readout plateau.

logical vector magnitude30.0 mT/m
safe logical maximum40.0 mT/m
limiting physical channelGX · 75%

Ideal per-axis amplitude example. Real systems also constrain slew rate, duty cycle, peripheral nerve stimulation, and vector-dependent safety limits.

04 / GRADIENT MOMENT & SPIN PHASE

The area under G
becomes phase.

Gradient amplitude alone does not set a k-space coordinate. Its signed time integral does. Build one lobe, add an opposite rewinder, and watch a three-dimensional spin ensemble wind into a phase pattern—or return coherently to k = 0.

ROTATING FRAME / IDEAL LINEAR GRADIENT20 mm UNIFORM SPIN COLUMN
Gradient axis
Three-dimensional spin phase ensemble
transverse magnetization vectors equal-phase planes
GX MOMENT kx = +170.3 m⁻¹ Phase is drawn in the transverse x–y plane; B₀ points along physical z.

Drag to orbit · scroll to zoom

ZEROTH GRADIENT MOMENT

M0(t) = ∫0t G(t′) dt′

POSITION IN K-SPACE

k(t) = γ̄ M0(t)

SPIN PHASE

φ(r,t) = 2π k(t) · r

Signed gradient area and accumulated kUNBALANCED
G(t)k(t) encode rewinder 0% +170.3 m⁻¹
+10.0 mT/m −300+30 mT/m
0.40 ms 00.601.20 ms
0% none100% · k = 0120%
COHERENT SIGNAL FROM THE COLUMN8.9%

The positive Gx area moves the sample to +kx. Spins separated along x retain different phases after the lobe turns off, so their vector sum is small.

net M₀+4.00 mT·ms/m
k coordinate+170.3 m⁻¹
phase across 20 mm+3.41 turns
01

Area, not height

A weak gradient held longer can create the same phase slope and k-space displacement as a short, strong gradient.

02

Polarity sets direction

Changing the sign of G reverses the phase ramp and moves to the opposite side of k-space along the selected logical axis.

03

Balanced area refocuses

For stationary spins in this ideal model, an equal opposite lobe cancels M₀. The phase ramp unwinds and the coherent signal returns at k = 0.

05 / LIVE CARTESIAN ACQUISITION

Build an image
from zero samples.

This is an interactive acquisition—not a prerecorded video and not an image being uncovered. Start with an intentionally empty reconstruction; each TR adds complex Fourier data from the whole slice, and the image is recalculated.

Δk · EXPLAIN EVERY WORD BEFORE USING THE FORMULA

Two numbered phase-pattern labels—and the exact numerical step between them.

A k-space “address” is not a place in the patient and has no width. It is a number attached to one measured complex sample. That number states how many gradient-created phase cycles occur per metre across the object for that sample.

“ADDRESS”A coordinate number such as kx = 9.10 m⁻¹.

It labels the phase pattern used when one whole-object I/Q sample was measured. It does not select a body point.

“NEIGHBOURING”Consecutive entries on the planned sampling list.

For example, 9.10 and 13.65 m⁻¹ are neighbors when no planned address lies between them.

“GAP”Subtraction on a number line—not empty physical space.

13.65 − 9.10 = 4.55 m⁻¹. An address itself has no size; Δk is only the numerical separation.

“CYCLES PER METRE”Cycles of relative spin phase across distance.

One cycle is a full 360° change in the gradient-created phase pattern, not one RF carrier oscillation and not a proton orbit.

Number line + phase patterns + repeating image periodΔk 4.55 m⁻¹ · FOV 220 mm
Interactive explanation of neighboring k-space addresses and their spacing Two neighboring numerical k-space coordinates differ by delta k. Their phase patterns differ by one complete cycle across the reciprocal field of view, and that field of view is the repetition distance of the reconstructed image. 01 · PLANNED kx NUMBER LINE · UNIT = CYCLES OF PHASE PER METRE subtract → Δk = 4.55 m⁻¹ 02 · WHAT THOSE TWO LABELS TELL THE SPIN-PHASE PATTERN TO DO ADDRESS A · kx = +9.10 m⁻¹2.00 turns across this FOV ADDRESS B · kx = +13.65 m⁻¹3.00 turns across this FOV +1 turn 03 · RECIPROCAL RESULT · THE RECONSTRUCTED OBJECT REPEATS EVERY FOV = 1/Δk one repetition = 220 mm FOV
FOURIER SAMPLE MICROSCOPEEXACT 64 × 64 COMPLEX DFT · 220 mm FOV
Every spin’s complex contributionUNIFORM PHASE AT k = 0

Color is the phase of ρeff(x,y)e−i2πk·r. Here ρeff means the local echo signal after the sequence and receive-coil factors explained in the rho lab. The receiver adds every colored contribution into one complex number.

The sample’s k-space addresskx 0 · ky 0
kx
ky
LOG |S(k)|LOWHIGH

The background is the phantom’s exact Fourier spectrum. The coral ring marks the single receiver sample inspected at left.

ONE RECEIVER SAMPLE

S(kx,ky) = ∫∫ ρeff(x,y)e−i2π(kxx+kyy) dxdy

one k-space coordinate is a weighted sum from the entire excited slice
0 Δk · 0.0 m⁻¹ −32 Δkcenter+31 Δk
0 Δk · 0.0 m⁻¹ −32 Δkcenter+31 Δk
COHERENT DC SAMPLE

At k = 0 the encoding phase is identical everywhere. All positive spin density adds coherently, producing the large center coefficient that represents the object’s average signal—not a center pixel.

|S(k)| / |S(0)|100.00%
sample phase0.0°
normalized complex sample+1.000 + i0.000
required gradient momentMx 0.000 · My 0.000
phase cycles / FOVX 0 · Y 0
encoded wavelengthuniform phase
GY PREPHASEky = 0 set · Gx / ADC idle
READY · SELECT k OR SWEEP
ONE-TR CONDUCTOR / SPOILED GREEVENT-EXPANDED TIME AXIS · EXACT PHYSICAL READOUT VALUES
RF, gradients, receiver, and signalTE 6 ms · TR 30 ms

The event block is expanded so short RF and gradient operations remain visible; the broken segment compresses idle recovery before the next RF pulse.

Integrated gradient pathkx 0.0 · ky 0.0 m⁻¹

Violet is unsampled prephasing. Mint is the portion stored by the ADC. The yellow ring is kx = 0—the echo center on this ky row.

RF + GZ · SLICE EXCITATION

BODY TX ON · RX ARRAY DETUNED

+GZ · RF BAND SELECTS Z

The body transmit coil and slice-select gradient act together first: RF creates transverse magnetization only in the frequency-matched slab. No in-plane sample has been recorded yet.

TIME IN TR0.00 / 30 ms
K-SPACE COORDINATEkx 0.0 · ky 0.0 m⁻¹
ADC STATEOFF · 0 / 64 SAMPLES
RECEIVE MAGNITUDE0.0% · BEFORE ECHO
THE ACQUISITION CHAIN

k(t) = γ̄∫G(τ)dτ S(k) −1

gradients choose the Fourier address; the ADC stores the complex receiver voltage only while its gate is open
220 mm 160sets Δk + pixel320 mm
128 kHz 64receiver bandwidth256 kHz
−16 Δk −32Gy moment+31
6 ms 5RF center → echo25 ms
30 ms 15next RF + scan time100 ms
55 ms 20illustrative tissue160 ms
REFERENCE PRESCRIPTION

A 220 mm FOV sampled at 64 readout points produces 3.44 mm pixels and Δk = 4.55 m⁻¹. Scrub the timeline or change a parameter to see the dependent quantities move together.

Δk = 1 / FOV4.55 m⁻¹
nominal kx,max145.5 m⁻¹
ideal Δx = FOV / 643.44 mm
ADC dwell / window7.81 µs / 0.50 ms
readout Gx13.7 mT/m
Gy moment−1.708 mT·ms/m
signal magnitude left at echo by T₂*89.7%
64-line scan proxy1.92 s

Ideal one-line-per-TR Cartesian GRE model with 64 complex samples and no ramp sampling. “Total sample rate” is used here because vendor bandwidth displays may instead report Hz/pixel. Here T₂* controls only the slow decay of transverse signal magnitude with time; use the TE/TR lab above for the full spoiled-GRE steady state.

0%
0 / 64 0 / 64 ADC SAMPLES
SEQUENCE / GRE-CARTESIANINTERACTIVE · 0 / 4096 SAMPLES
BEFORE THE FIRST TR

Frequency is not amplitude.

Open the definitions below, then press Run. This explanation will follow RF transmit, gradients, receive sampling, and reconstruction through each repetition.

RF carrier frequency≈ 127.73 MHz at 3.0 T+

How fast the transmit field oscillates and the received voltage alternates. It is tuned near the ¹H Larmor resonance. It may be offset or shaped to select a slice, but it is not the waveform height.

RF amplitude · B₁⁺µT · slow strength outline+

How strong the transmit field is. Together with pulse duration it sets flip angle. The drawn outline is simply B₁⁺ strength versus time; it cannot display the tens to hundreds of millions of carrier cycles per second at this scale.

Gradient · Gx, Gy, GzmT/m · field slope+

Not a radio wave. A gradient slightly changes local Larmor frequency with position. Its signed area sets retained phase and the k address; while it remains on, its amplitude sets how many inverse metres that numerical address changes per second.

Received signalcomplex voltage · I + iQ+

The coil detects a tiny RF voltage near the carrier. The receiver removes that fast carrier and stores a complex sample: magnitude says how much coherent signal arrived and phase preserves spatial encoding.

Image brightness|inverse Fourier transform|+

Not raw RF amplitude and not a k-space location. Reconstruction combines every acquired complex sample, and magnitude display maps the resulting voxel signal to brightness.

START · 0 OF 4096 COMPLEX SAMPLES

No acquired data means no MRI image.

The reconstruction canvas is intentionally empty. Press Run or drag the bottom progress slider; it controls the simulated acquisition and recalculates the image from only the samples acquired so far.

  1. 00No datano image yet
  2. 01Near k = 0broad shape + contrast
  3. 02Larger |k|edges + fine detail
  4. 03All rowscomplete ideal model

THE RULEA k-space sample does not paint one image pixel. Every acquired complex coefficient changes the calculation of every reconstructed pixel.

Pulse sequence / one TRTR 01 / 64
RFGzGyGxADC α slice select phase +31 readout 64 complex samples
ExciteSelect zSet kyTraverse kx
Raw signal / k-spacekx −32 · ky +31
kx
ky

Each acquired cell stores S = I + iQ. Its displayed brightness is L = ln(1 + √(I² + Q²)); this compresses the preview only. Position in k-space is spatial frequency—not a location in the head.

Fourier reconstruction0% DATA
0 / 4096 COMPLEX SAMPLES NO MRI IMAGE YET Run the acquisition or drag the progress slider.
AR
Δx = 3.4 mm

0 samples: the canvas is deliberately empty because no image can be reconstructed yet.

HIGHER-RESOLUTION FOURIER LAB · 64² → 128² → 256² COMPLEX SAMPLES

Keep the reconstruction.
Add detail and tissue contrast.

This is a second, independent reconstruction built from a 256 × 256 digital teaching phantom. Matrix changes how far the sampled grid reaches into spatial frequency at a fixed 220 mm field of view. TE and TR change each tissue’s signal before encoding. They are separate causes, so the controls never pretend that contrast timing changes pixel size.

01 · MATRIXChanges detail support

At fixed FOV and Δk, more samples extend to larger |k| and make smaller pixels. It also requires more phase-encoding repetitions in this one-line-per-TR model.

02 · TEChanges T₂ survival

Longer echo time waits longer before k-space center is sampled. Short-T₂ tissues lose more coherent spin-echo signal than long-T₂ fluid.

03 · TRChanges T₁ recovery + time

Longer repetition time lets more longitudinal magnetization recover before the next RF pulse and lengthens this simplified scan-time estimate.

04 · FOURIEREvery coefficient is global

One I + iQ sample is one whole-object spatial pattern. The inverse Fourier transform combines all acquired coefficients into every reconstructed pixel.

ADVANCED COMPLETE · 65,536 OF 65,536 COMPLEX SAMPLES

All sampled spatial-frequency patterns now contribute at 256 × 256.

The full matrix contains four times as many samples along each axis—and sixteen times as many complex coefficients—as the 64 × 64 beginner reconstruction. Every stored coefficient contributes to every output pixel; no photograph is uncovered from underneath.

01 · OBJECT-SPACE INPUT MODELTE/TR-weighted tissue signal · full 256 reference grid

This first panel is a known digital input used to test the math—not an image secretly revealed during acquisition. Its gray levels use the spin-echo signal equation for five illustrative tissue classes.

02 · ACQUIRED COMPLEX DATA256 × 256 support · log |I + iQ| preview

Mint pixels are acquired I + iQ coefficients. Coral rows are planned but not yet acquired; the dark outer area lies beyond the selected matrix. Log magnitude makes weak coefficients visible but does not discard their stored phase.

03 · INVERSE FOURIER RESULT0.86 mm pixels · complete ideal matrix
nominal acquired interval 0.86 mm

At full 256 × 256 support the finest digital phantom targets are better separated. This is ideal sampling detail, not guaranteed clinical diagnostic resolution or a patient image.

DETAIL MICROSCOPE · THE SAME PHYSICAL SQUARE ON BOTH IMAGES

Do not hunt for the extra detail.
Put it side by side.

Drag either mint square with a mouse or finger. Both squares stay locked to the same place. The left magnifier shows the declared digital source; the right magnifier shows what the currently acquired I + iQ data can reconstruct there. Arrow keys move a focused square; Shift + arrow moves it farther.

A · KNOWN SOURCE CROPWhat the numerical phantom actually contains

This is not scanner output. It is the answer key supplied to the Fourier calculation, magnified with hard square display bins so its tiny targets are visible.

SAME WINDOW · LIVE MEASUREMENT LEDGER
34.38 mm square near the lower-left teaching targets
PHYSICAL WINDOW WIDTH34.38 × 34.38 mmsame on source and result
ACQUIRED INTERVAL0.86 mm220 mm FOV ÷ matrix
INTERVALS ACROSS WINDOW40.0 cellsyellow grid on the result
SCREEN DISPLAY BIN0.86 mmfixed 256-bin output grid
CLICK FORMULA · CHANGE N IN THE LIVE EXAMPLE nominal acquired interval = FOV ÷ N

220 mm ÷ 256 acquired samples = 0.86 mm. This number says how the sampled width is divided; it does not prove that a 0.86 mm object is visibly resolved.

The yellow lines mark the 256 × 256 acquisition intervals. At this setting one acquired interval and one displayed bin have the same width.

B · CURRENT FOURIER CROPWhat the retained complex samples support

All 65,536 complex samples are present. Compare the circle gaps and line edges with the known source; similarity here follows the ideal sampled Fourier data, not a hidden photograph.

01 · SMALLER SCREEN SQUARES ARE NOT AUTOMATICALLY MORE INFORMATION
Display grid ≠ acquired detail ≠ measured sharpness

This teaching result always uses 256 display bins across 220 mm. With a 64 matrix, each acquired interval spans four display bins in each direction; the extra in-between gray values are calculated from the same 64 × 64 Fourier data. They make a smoother-looking screen, but cannot invent missing outer-k-space patterns.

02 · WHAT “ACTUAL SHARPNESS” WOULD REQUIRE
A point response must stay narrow enough

An ideal mathematical point becomes the system’s point-spread function, or PSF. Finite k-space, filters, gradient errors, relaxation during readout, motion, off-resonance, and reconstruction can broaden that response. Two tiny objects blur together when their broadened responses overlap too much—even if the screen pixels are smaller.

03 · WHY REAL HIGH-RESOLUTION SCANS CAN LOOK GRAINY
This lab has zero random receiver noise

Its signal-to-noise ratio is deliberately not estimated. In a scanner, smaller voxels usually contain less signal, and weak outer-k-space measurements may compete with random coil and receiver noise. More matrix is useful only when signal, scan time, coils, motion control, bandwidth, gradients, and reconstruction preserve the added information.

Near k = 0

Contrast & broad shape

Slow spatial variation. High signal energy. Acquired at the echo center.

Large |k|

Edges & fine detail

Rapid spatial variation. Extending farther raises the ideal resolution limit.

Sample spacing Δk

Field of view

Closer k-space samples encode a wider unaliased FOV: FOV = 1 / Δk.

06 / TRAJECTORY STUDIO

Gradients draw
the path.

Here is a spatial-phase address, not a physical position or a moving particle. A gradient changes that address over time: hold one component constant and the address follows a straight line; reverse it and the address turns back; vary two components together and the address can spiral. Compare five encoding strategies built from that rule.

WHAT DOES k MEAN?

Count full turns of relative spin phase across distance.

k = 100 m⁻¹ means the gradient-created phase pattern completes 100 turns per metre. Therefore two fixed positions 10 mm apart differ by one full 360° turn at that instant.

The marker is the current data address where ADC stores a whole-object complex sample. It is not a proton, voxel, anatomical location, RF carrier cycle, or signal amplitude.

WHY A GRADIENT “MOVES” k dk/dt = γ̄G(t)

dk/dt is the rate at which the spatial-phase address changes, in m⁻¹/s. G(t) is the field slope in T/m. Positive G moves toward +k, negative G toward −k, and G = 0 holds the address still.

WHY CARE · The visited extent sets potential detail, spacing sets FOV, and a wrong address from delay or miscalibration creates blur, ghosts, or geometric distortion.

TRAJECTORY / IDEAL COMMANDREADY
Interactive k-space trajectory
commanded k(t) delayed response
CURRENT SAMPLE k = (−1.00, −1.00, 0.00)

Drag to orbit · scroll to zoom

MULTI-SHOT · RECTILINEAR

One echo, one row.

A prephaser sets the negative-kx starting address. The readout gradient traverses one constant-ky line while ADC samples. On the following repetition, a different signed phase-gradient area creates a different retained phase ramp and sets the next ky value.

THE TRAJECTORY LAWG is a field slope. Its direction chooses which k coordinate changes; its magnitude sets the address-change rate in m⁻¹/s.

dkdt= γ̄ G(t)

Representative gradient commandSHOT 01 / 15
GxGyGz ADC
0%
0 µs ideal10 µs20 µs

i Delayed or filtered gradient response means the scanner may sample somewhere other than the commanded coordinate. Non-Cartesian reconstruction therefore often uses a measured trajectory.

coverage / excitation1 k-space row
center crossingsonce per echo
reconstructiondirect FFT
G(t)

Gradient amplitude

Controls how rapidly the numerical k address changes. A stronger readout gradient covers more spatial-frequency address per unit time; nothing physically flies through the patient.

dG/dt

Slew rate

Limits how sharply a path can turn. Fast switching also drives acoustic noise and peripheral nerve-stimulation constraints.

ADC(t)

Sampling window

The path may move while the receiver is off. Only coordinates visited during ADC become acquired data samples.

07 / RESOLUTION LAB

Shape the voxel
in all three axes.

Resolution is not a “megapixel” setting. It follows from field of view, sample count, and how far the acquisition reaches in k-space. Smaller voxels usually cost signal-to-noise, scan time, or both.

3D ENCODED VOLUME1.72 × 1.72 × 5.00 mm

Drag to orbit · highlighted cell is one voxel

Protocol builder
XReadoutfrequency
220 mm
128
1.72 mm
YPhaseencoded
220 mm
128
1.72 mm
ZSliceRF selected
2.0 kHz
9.4 mT/m
5.00 mm
voxel volume14.77 mm³relative SNR proxy 1.00×
k-space extent±291 / ±291 m⁻¹Z uses slice profile
minimum phase encodes128≈ 25.6 s at TR 200 ms
REFERENCE PRESCRIPTIONWhat changed—and what it costs

This 220 mm, 128 × 128 reference produces 1.72 mm in-plane sampling. Move any control to see its direct consequence for detail, signal, and encoding time.

in-plane pixel area 1.00× signal proxy 1.00× phase-time proxy 1.00×
PIXEL / VOXEL SIZE

Δx = FOVx / Nx ≈ 1 / (2kx,max)

FIELD OF VIEW

FOVx = 1 / Δkx

2D SLICE THICKNESS

Δz = BWRF / (γ̄ |Gz|)

SLICE MICROSCOPE / RF + GZIDEAL LINEAR FIELD · ¹H
Gz polarity
Three-dimensional RF-selected slice
LOWER FREQUENCYHIGHER FREQUENCY
RF-SELECTED SLAB 4.70 mm center z = 0.0 mm · axial plane

Drag to orbit · highlighted rings fall inside the transmit passband

FREQUENCY ADDRESS

Δf(z) = γ̄ Gz z

SELECTED THICKNESS

Δz = BWRF / (γ̄ |Gz|)

Frequency-to-position map+GZ · CENTERED RF
Δfz−1200+120 mm RF −1.0…+1.0 kHz selected z = 0.0 mm
10.0 mT/m 41730 mT/m
2.0 kHz 0.53.256.0 kHz
0.0 kHz −20on-resonance+20 kHz
CENTERED AXIAL SLICE

Gz maps position to frequency. The 2.0 kHz RF passband intersects that line around z = 0, exciting an ideal 4.70 mm slab.

frequency slope0.426 kHz/mm
RF passband−1.0 … +1.0 kHz
slice center0.0 mm
ideal thickness4.70 mm
BWRF

Thicker slab

A wider transmit band matches a larger interval of positions on the same frequency slope.

|Gz| ↑

Thinner slab

A steeper frequency slope maps the same RF bandwidth onto a narrower spatial interval.

fRF

Move the slice

Changing the RF center frequency moves the intersection without changing the ideal thickness.

08 / SAMPLING & POINT RESPONSE

Change k-space.
The image answers.

Spatial resolution, field of view, ringing, and wraparound are different consequences of how k-space is sampled. Apply the operations directly to the complex data and inspect both the reconstructed image and the point-spread function that produced it.

FOURIER LAB / COMPLEX DATA64 × 64 PHANTOM
Apply along
Sampling prescription
CUTOFF / RESOLUTION

Δx ≈ 1 / (2kx,max)

SPACING / FIELD OF VIEW

FOVx = 1 / Δkx

100% kmax 25% · blurred60%100% · full
Uniform sampling interval
k-space weighting
FULL REFERENCE

All encoded frequencies are retained at the native Δk. The PSF approaches one image pixel and no coherent wrap replicas are introduced.

nominal Δx × Δy3.44 × 3.44 mm
effective FOV220 × 220 mm
PSF FWHM1.0 × 1.0 px
retained samples4,096 / 4,096
Applied sampling mask100% RETAINED
kykx

Mint points contribute to reconstruction. Coral lines are inside the selected extent but skipped by uniform undersampling.

Image-space PSFℱ⁻¹{MASK}
positivenegative

The central lobe sets effective resolution; sidelobes produce ringing or displaced wrap replicas.

Resulting reconstructionREFERENCE
AR

Full k-space extent and native spacing preserve the simulated image’s available detail and field of view.

01

Truncate extent

Removing outer k-space lowers the spatial-frequency cutoff. Image space is convolved with a broader sinc-like response: detail softens and sharp boundaries ring.

02

Increase spacing

Keeping every Rth point multiplies the effective Δk by R and reduces the encoded FOV by R. Repeated PSF peaks fold distant anatomy into the displayed FOV.

03

Apodize

A Hann or Hamming taper suppresses PSF sidelobes and Gibbs ringing, but broadens the central lobe. A calmer boundary is purchased with effective resolution.

09 / THE COMPLETE CHAIN

From current waveform to one image voxel.

  1. 01

    Gradient current

    Amplifiers drive X, Y, Z coil windings. Current and geometry create a controlled field slope in T/m.

    I(t) → G(t)
  2. 02

    Spin phase

    The local field changes Larmor frequency. Accumulated phase records the gradient’s area through time.

    φ(r,t) = γ r·∫G dt
  3. 03

    k-space sample

    The receiver sums every transverse spin at the current spatial-frequency coordinate.

    k(t) = γ̄∫G dt
  4. 04

    Image estimate

    An inverse Fourier transform separates the superposed spatial frequencies into locations.

    ρ̂(r) = ℱ⁻¹{S(k)}

REFERENCE DESK

Continue into the physics.

FORMULA EXPLAINER

Formula explanation

Dotted technical words open their age-10 dictionary entry. Close that entry—or press Escape—to return to this exact formula.

READ THE EQUATION IN WORDS

LIVE PHYSICS PICTURE

    WHAT IS ACTUALLY MEASURED?

    Separate commands, physical quantities, and calculated results

    Symbols & units

    If one quantity changes

    UNIT DECODER

    Every abbreviation, prefix, and conversion

    Capitalization is part of the unit: M means mega (10⁶), while m can mean milli (10⁻³) or metre depending on its position.

    WORKED WHAT-IF

    Change one number

    0

    OUTPUT

    PHYSICS CONSEQUENCE

    CLINICAL / IMAGE CONSEQUENCE

    VISUAL GUIDE

    What this view shows

    Dotted technical words open their age-10 dictionary entry. Close that entry—or press Escape—to return to this exact visual key.

      TRY IT

      Controls that reveal the relationship

        MRI LANGUAGE LENS · ABBREVIATION & UNIT DICTIONARY

        Decode every symbol.

        All 123 entries first show an age-10 meaning, symbol decode, physical/calculated/displayed status, exact unit, why MRI needs it, what fails without it, and what more/less/negative does—then a labeled visual and deeper physics. Technical words inside an entry are clickable too. Back and Forward preserve a multi-term reading trail.

        ENCODING
        01 / 01

        Readout

        Gread + ADC
        START HERE · NO MRI KNOWLEDGE ASSUMED

        Explain it as if I am 10

        DECODE THE NAME / SYMBOL

        WHAT KIND OF THING IS IT?

        MEASURED OR WRITTEN IN

        WHY MRI NEEDS IT

        WITHOUT IT—or IF IT IS WRONG

        WHEN IT BECOMES MORE, LESS, OR NEGATIVE

        Reading rule: a number is meaningful only when its unit, reference, direction, and held-fixed conditions are stated. This entry states all four whenever they apply.

        ONE LEVEL DEEPER

        VISUAL MODEL · EVERY OBJECT EXPLAINED

        SELECTED OBJECT · 01 OF 04

        PHYSICS / SCANNER CONNECTION
          WHAT THE SCANNER DOES

          WHAT CHANGES IN DATA / IMAGE

          WHY YOU CARE

          DO NOT CONFUSE IT WITH

          See it operate in the site